Faster than traditional methods of mobile phase preparation, saving time in the laboratory.
-
Enhanced laboratory safety.
-
No need to purchase expensive degassing equipment.
-Whatman in-line filter/degassers (IFD) connect directly into an HPLC line to simultaneously filter and degas the mobile phase as it is being used. The Solvent IFD is used with organically based HPLC mobile phases.-Solvent IFD has a 0.2 μm high-flow polypropylene membrane for mobile phases containing organic solvents. They have a polypropylene housing, the circumference of which is sealed by a security ring, fittings to accommodate 1/16”-1/8” tubing, and an air vent on the inlet with luer lock cap to enable priming.-
The in-line filters work on the principle of ''bubble point'' – the point of pressure at which gases will pass through a wet membrane. If pressure is maintained below the bubble point, the gas will not pass through the membrane and is trapped by the particular filter device.