Hellma® calibration standard set UV003 for checking the photometer:

Checking the wavelength accuracy with UV5 holmium oxide liquid filter, calibrated at wavelengths 241, 287, 361, 451, 485, 536, 640 nm, slit width 1 nm, measured against UV14 perchloric acid (blank).

Checking the photometric accuracy with UV60 potassium dichromate 60 mg/l, calibrated at wavelengths 235, 257, 313, 350 nm, slit width 2 nm; UV600 potassium dichromate 600 mg/l, calibrated at wavelength 430 nm, slit width 2 nm; both filters measured against UV14 perchloric acid (blank).

Checking stray light with UV1 potassium chloride calibrated at 198 nm (cut-off approx. 200 nm), slit width 2 nm, measured against UV12 ultrapure water (blank).

Checking the spectral resolution with UV6 toluene in hexane and UV9 hexane reference filter (blank), calibrated at wavelengths 265 - 270 nm, slit widths 0.5, 1.0, 1.5, 2.0, 3.0 nm.

All filters with serial number, packed in high-quality storage box with DAkkS calibration certificates, for UV6/UV9 Hellma Analytics calibration certificate. Note: Hellma Analytics calibration certificate (factory calibration. not covered by accreditation!)

Hellma® calibration standards are manufactured 100% in Germany.

Compliant with regulations
Hellma® calibration standards comply with the specifications of quality management
systems and pharmacopoeias for the highest quality requirements and international
comparability of the measurement results.

DAkkS calibration certificate*
Hellma® calibration standards are manufactured in the Hellma Analytics calibration laboratory accredited by the DAkkS* in accordance with DIN EN ISO/IEC 17025 and are supplied with a DAkkS calibration certificate*.

Traceability
All measurement results of Hellma® calibration standards can be traced back to calibration standards measured by NIST (photometric accuracy) or PTB ("Physikalisch-Technische Bundesanstalt") (wavelength accuracy).

Resource-saving and sustainable
Hellma® calibration standards can be reused for years without any problems and remain in use for a long time. There is no need to repeatedly prepare and produce calibration solutions, which saves resources, the environment and the budget. A sustainable and environmentally friendly alternative!

30-year guarantee
We trust in our quality and you can rely on accurate measurement results!
With regular recalibration - at least every two years - in the Hellma Analytics calibration laboratory, you receive a 30-year manufacturer's guarantee on all Hellma® calibration standards.

More Information

  • Additional
Brand Hellma
Type Liquid Filter

You might also be interested in the following products

  1. Product Image of Hellma® calibration standard UV1 with potassium chloride for checking stray light (198 nm), Dimensions: 48,5 x 12,5 x 12,5 mm
    Hellma® calibration standard UV1 with potassium chloride for checking stray light (198 nm), Dimensions: 48,5 x 12,5 x 12,5 mm
    Item number: HL667001
    Brand: Hellma
    $1,129.00