SPME (Solid Phase Micro Extraction) is a procedure of sampling. Later the samples can be analyzed via GC or HPLC. The sampling by SPME, which no solvent, was developed and patented by the chemist Janusz Pawlizyn. Marketing is carried out by the only licensee SUPELCO, which was part of the Sigma-Aldrich group and is now part of Merck.
Samples which are to be analyzed by gas chromatography are usually filled in so-called autosampler vials, which are sealed with a crimp lid. For sampling, an SPME needle is pierced through the septum into the sample vial. An SPME needle looks similar to a normal cannula because it is also made of stainless steel. However, this does not serve to remove or inject fluids, but helps in penetrating the septum. There is a fused silica (fused silica) or glass fiber (StableFlex) in the needle, coated on the end. This coated fiber absorbs analyte molecules from the sample. The sample can either be taken directly from the liquid phase or from the gas phase above the liquid, the so-called head-space. If the sample is to be taken out of the gas phase, a volatility of the analyte has to be given, so that enrichment from the gas phase is possible. More information about Headspace analysis can be found here.
Desorption
Following the extraction of the analyte from the liquid sample, is the desorption in order to release the analytes from the coated fiber of the SPME needle. Different methods are suitable depending on the choice of the desired analytical method. If sample is getting analyzed by of gas chromatography, the desorption can be carried out without further intermediate steps within the sample inlet of the gas chromatograph. In this case, the so called thermal desorption is used, using the high temperature within the chromatograph to dissolve the absorbed analyte molecules from the fiber and to convert them into the gas phase. The steps of thermal desorption of the analytes and transfer into the gas phase for the gas-chromatographic separation are connected and thus analyte loss is avoided or at least minimized. The temperatures of the Inlet depend on the respective temperature during the chromatographic separation, but usually are between 100 and 300 ° C.
The absorbed analyte molecules can also be dissolved from the fiber with the aid of a solvent. This method is used if later analysis is to be carried out by HPLC (high pressure liquid chromatography).
Coating materials
Through the use of different coating materials, substance groups can be selectively absorbed. The commercially available adsorbing materials which differ greatly in their polarity include: carbowax, polydivinylbenzene, polyacrylate, carboxes and PDMS (polydimethylsiloxane). Depending on the analyte, combinations of different layer thicknesses and adsorbing materials are used.
The following table lists the fibers recommended by Merck for various types of analytes (MW = molar mass):
Analyt Type
Recommended Fiber
Gases and low molecular weight compounds (MW 30-225)
75 µm/85 µm Carboxen/polydimethylsiloxane
Volatiles (MW 60-275)
100 µm polydimethylsiloxane
Volatiles, amines and nitro-aromatic compounds (MW 50-300)
65 µm polydimethylsiloxane/divinylbenzene
Polar semi-volatiles (MW 80-300)
85 µm polyacrylate
Non-polar high molecular weight compounds (MW 125-600)
7 µm polydimethylsiloxane
Non-polar semi-volatiles (MW 80-500)
30 µm polydimethylsiloxane
Alcohols and polar compounds (MW 40-275)
60 µm Carbowax (PEG)
Flavor compounds: volatiles and semi-volatiles, C3-C20 (MW 40-275)
50/30 µm divinylbenzene/Carboxen on polydimethylsiloxane on a StableFlex fiber
Trace compound analysis (MW 40-275)
50/30 µm divinylbenzene/Carboxen on polydimethylsiloxane on a 2 cm StableFlex fiber
Amines and polar compounds (HPLC use only)
60 µm polydimethylsiloxane/divinylbenzene
Advantages of SPME:
Solvent-free micro-extraction technology for almost all samples and matrices